Appendix B - EMC test levels.
Test standard Description
Class/Group/Immunity
test level
CISPR11:2009+A1:2010
Radiated emissions Class A Group 2
CISPR11:2009+A1:2010
Conducted
emissions
Class A Group 2
IEC/EN 61000-4-2
Immunity from
electrostatic
discharge
±15kV air, ±8kV contact
IEC/EN 61000-4-3
Radiated RF
immunity
3V/m
IEC/EN 61000-4-3
Radiated immunity
from intentional
transmitters
28V/m maximum
IEC/EN 61000-4-4
Immunity from
electrical fast
transients and bursts
±2kV AC supply line,
±1kV signal lines
IEC/EN 61000-4-5
Surge immunity on
AC supply
±2kV common mode,
±1kV differential mode
IEC/EN 61000-4-6
Conducted RF
immunity
3V rms 150kHz > 80MHz,
6V rms ISM and amateur
bands
IEC/EN 61000-4-11
Immunity to voltage
dips, short
interruptions and
voltage variations
10ms > 5s dip/interruption
time