OM265EN issue 16
39
Appendix A - Technical Specification
General
Power Input 100-240 Vac 50/60 Hz
Classification (EN60601-1) Class 1, Type BF
Mains Fuses 2 x T10A 250V (5 x 20 mm)
Size (height x width x depth) 940 x 470 x 470 mm
Weight 38 kg (excluding electrodes)
Treatment Programs 10 user-defined set-ups
Shortwave
Frequency 27.12 MHz
Maximum Output Power 400 W in continuous mode
1000 W peak in pulsed modes
Modes Continuous, 3 in 3, 2 in 3 and 1 in 3
Pulse Frequency 5 - 800 Hz
Pulse Width 20 - 400 µs
Tuning Automatic
Treatment Timer 0 to 30 minutes
The Megapulse Senior 265 is designed to operate from any 50/60 Hz
single phase supply between 100 and 240 Vac capable of supplying
1 kVA. Connection is via an IEC socket at the rear of the unit.
All information on model, serial number, and month/year of
manufacture is located on the rear panel.
Each Megapulse Senior 265 is supplied with a detachable mains
cable, spare fuses, a pair of 100mm capacitive electrodes, output
tester and this manual.
WARNING – Class 1 equipment - to prevent electric shock connect
to protective earth.