EasyManua.ls Logo

EMS PIEZON 250 - Cleaning

EMS PIEZON 250
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
Reusable Instruments and Accessories for the Piezon
150 and Piezon 250
Description Description
INSTRUMENTS Instrument RE2
Intrument A Instrument PI
Instrument C Soft Instrument
Instrument B Instrument RT4
Instrument H Cap SP
Instrument D File ISO 15
Instrument E File ISO 20
Instrument F File ISO 25
Instrument G File ISO 30
Endochuck 120° File ISO 35
Instrument P Flat wrench
Endochuck 90° COMBITORQUE
Instrument SP Combitorque
Instrument PS HANDPIECE HOLDER
Instrument PL1 Handpiece Holder diam 16
Instrument PL2 HANDPIECE
Instrument PL3 Handpiece Piezon
Instrument PL4 Handpiece Piezon LED
Instrument PL5
Instrument PI
Endo-chuck 180°
Endochuck 120°
Instrument RT3
Recommended cleaning, packaging for sterilization, and sterilization procedures are provided
below.
Cleaning
Instruments, Handpiece Connector, Combitorque
Remove any externally attached soiling by brushing carefully with a soft surface brush or
cloth.
Remove the instruments from handpieces.
Soak all parts in the cleaning solution for, at a minimum, the time and the concentration
speci ed by the manufacturer of the cleaning agent (ENZOL, 2 %, 10 min).
Remove any externally attached soiling by brushing carefully with a soft surface brush or
cloth.
Place all parts an ultrasound bath lled with cleaning solution (ENZOL, 2 %).
Treat the parts with ultrasound for at least 3 minutes.
Brush the parts in the cleaning solution (ENZOL, 2 %) if any visible soiling is detected.
Submerge all components completely in deionized water.
Rinse through all product lumens with deionized water.
Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible
on the product.
Dry all components at room temperature.
Handpieces
Remove any externally attached soiling by brushing carefully with a soft surface brush or
cloth.
Remove the instruments from the handpieces. Remove the handpieces from the
connectors. Remove the gaskets, caps and other removable parts.
Soak all parts in the cleaning solution for, as a minimum, the time and the concentration
speci ed by the manufacturer of the cleaning agent (ENZOL, 2 %, 10 min).
Brush all lumens and surfaces with a suitable lumen / surface brush until all visible soiling
is removed.
Rinse through all product lumina (e.g. irrigation and aspiration connection) with the
cleaning solution (ENZOL, 2 %), at least 5 times in the ow direction (no back rinsing),
using a disposable syringe (min. volume 50 ml) applied to the nozzles of the product.
Submerge all components completely in deionized water.
Rinse through all product lumina (e.g. irrigation and aspiration connection) with deionized
water at least 3 times in the ow direction (no back rinsing), using a disposable syringe
(min. volume 50 ml) applied to the nozzles of the product.
Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible
on the product.
Dry all components at room temperature.