6.7 Evaporating Unit (Option)
WARNING
Explosion hazard!
Do not use per
-chloric acid.
No organic solvents (alcohols, hydrocarbons,
etc.) may be placed in the pressure vessels
or the microwave.
Highly reactive substances which spontaneously react at higher tem
peratures or in the
presence of oxidizing agents or which may be nitrated and are potentially explosive may
not be digested in the microwave. Examples of such substances include: explosives
(TNT, etc.), perchlorates, ether, pyrophoric substances, etc
.
WARNING
Always wear safety goggles and protective gloves!
Always wear safety glasses/goggles and protective gloves during work which may in-
volve contact with chemicals, that is, when working with acids, pressur
e vessels, during
cleaning, etc.
WARNING
Hot acid vapors are generated! Risk of acid burn!
Hot acid vapors are generated during evaporating. Take particular care to ensure that all
connections are secure and do not leak. In particular, the tube connection between the
microwave and the gas scrubbe
r must always be totally sealed.
Secure the tube appropriately to ensure that no acid vapors can escape in the direction
of surrounding individuals, even when the screw
-on connection is opened.
The microwave system, together with the gas scrubber, must be
set up under a vent
hood.
1 PFA gas exhaust
2 Connection for water cooling-outlet
3 Connection for water cooling-inlet
4 System for CX 100
5 System for PM 60 and PL 100