Type Mode de protection Alimentation
(bornes 1+ et 2-)
Circuit du
capteur (bornes
3 à 7)
Valeurs de raccordement max.
TMT71,
TMT72
Ex ia IIIC T85°C...T120°C Dc U
i
≤ 30 V
DC
I
i
≤ 100 mA
C
i
= négligeable
L
i
= négligeable
U
o
≤ 4,3 V
DC
I
o
≤ 4,8 mA
P
o
≤ 5,2 mW
Ex ia IIIC/IIIB/IIIA L
o
= 100 mH
C
o
= 18 µF
TMT71,
TMT72
Ex tc IIIC T85°C...T105°C Dc
Ex tc IIIC T105°C Dc
Ex tc IIIC T120°C Dc
1)
Ex nA IIC T6...T4 Gc
U
b
= 10 … 36 V
DC
Sortie : 4...20mA
1) Température de surface maximale admissible au boîtier pour les transmetteurs sans tête de raccordement fournie