Levelflex FMP56, FMP57
Endress+Hauser 11
Input
Measured variable
The measured variable is the distance between the reference point and the product surface.
Subject to the empty distance entered "E" the level is calculated.
Alternatively, the level can be converted into other variables (volume, mass) by means of
linearization (32 points).
Measuring range
The following table describes the media groups and the possible measuring range as a function of
the media group.
Levelflex FMP56
Media group DC (ε
r
) Typical bulk solids
Measuring range
bare metallic
rope probes
PA-coated
rope probes
1 1.4...1.6 plastic powder 12 m (39 ft)
1)
2 1.6...1.9 plastic granulate
white lime, special cement
sugar
12 m (39 ft) 12 m (39 ft)
3 1.9...2.5 portland cement, plaster 12 m (39 ft)
flour 12 m (39 ft)
4 2.5...4 grain, seeds 12 m (39 ft)
ground stones
sand
12 m (39 ft) 12 m (39 ft)
5 4...7 naturally moist (ground) stones, ores
salt
12 m (39 ft) 12 m (39 ft)
1) Restrictionss: for materials with a strong signal damping, e.g. ground material, wheat bran or silicic acid, the measuring range may be reduced.