Levelflex FMP56, FMP57 FOUNDATION Fieldbus Operating menu
Endress+Hauser 205
Level signal
Navigation  Diagnostics → Device check → Level signal
Prerequisite Device check has been performed.
Description Displays result of the device check for the level signal.
User interface Check not done
Check not OK
Check OK
Additional information For Level signal = Check not OK: Check the mounting position of the device and the
dielectric constant of the medium.
Launch signal
Navigation  Diagnostics → Device check → Launch signal
Prerequisite Device check has been performed.
Description Displays result of the display check for the launch signal.
User interface Check not done
Check not OK
Check OK
Additional information For Launch signal = Check not OK: Check the mounting position of the device. In non-
metallic vessels use a metal plate or a metal flange.