Operating menu Levelflex FMP56, FMP57 FOUNDATION Fieldbus
204 Endress+Hauser
17.4.8 "Device check" submenu
Navigation  Diagnostics → Device check
Start device check
Navigation  Diagnostics → Device check → Start dev. check
Description Start a device check.
Selection No
Yes
Factory setting No
Additional information In the case of a lost echo a device check can not be performed.
Result device check
Navigation  Diagnostics → Device check → Result dev.check
Description Displays the result of the device check.
Additional information Meaning of the display options
Installation ok
Measurement possible without restrictions.
Accuracy reduced
A measurement is possible. However, the measuring accuracy may be reduced due to the
signal amplitudes.
Measurement capability reduced
A measurement is currently possible. Howerver, there is the risk of an echo loss. Check
the mounting position of the device and the dielectric constant of the medium.
Check not done
No device check has been performed.
Last check time
Navigation  Diagnostics → Device check → Last check time
Description Displays the operating time at which the last device check has been performed.