Mounting Liquiphant FTL64
8 Endress+Hauser
~13 (0.5)
~5 (0.2)
~12.5 (0.49)
D
A B C
D
D
A0044069
 2 Typical switch points. Unit of measurement mm (in)
A Installation from above
B Installation from below
C Installation from the side
D Switch point
5.1.2 Take viscosity into consideration
Viscosity values
Low viscosity : < 2 000 mPa⋅s
High viscosity: > 2 000 to 10 000 mPa⋅s
Low viscosity
It is permitted to position the tuning fork within the installation socket.
> 25 (0.98)
> !50 (1.97)
A0042333
 3 Installation example for low-viscosity liquids. Unit of measurement mm (in)