Liquiphant M/S with FEL56 and Nivotester FTL325N
12 Endress+Hauser
A0022344-EN
1oo1 architecture (CONF 1)
Liquiphant (FEL56)
Settings 1) Density 0,7 / 0,5
2) MIN safety
Evaluated NAMUR transmitter
A0022180
Error current signal
> 2,1 mA
SIL SIL 2
HFT 0
SFF 78%
PFD
av
** 0,029 x 10
-2
Wiring scheme
A0022181
Ask the manufacturer in question for the NAMUR transmitter parameters
relevant to safety.
Recurrent test
e.g. lowering the level
** TI (test interval) = annual