Commissioning Liquistation CSF34
56 Endress+Hauser
7.3.4 Basic setup
Making basic settings
1. Switch to the Setup/Basic setup menu.
Make the following settings.
2. Device tag: Give your device any name of your choice (max. 32 characters).
3. Set date: Correct the set date if necessary.
4. Set time: Correct the set time if necessary.
5. Number of bottles: Correct the set number of bottles if necessary.
6. Bottle volume: Correct the set bottle volume if necessary.
For quick commissioning, you can ignore the additional settings for outputs etc. You
can make these settings later in the specific menus.
7. To return to the display overview: press the soft key for ESC for at least one second.
Your sampler now works with your basic settings. The sensors connected use the
factory settings of the sensor type in question and the individual calibration settings
that were last saved.
If you wish to configure your most important input and output parameters in the Basic setup :
‣
Configure the current inputs, relays, limit switches, cleaning cycles and device diagnostics
with the following submenus.
7.3.5 Sampling programs
Difference between program types
The following box provides an overview of the differences between the Basic, Standard and
Advanced program types.
Basic (1 sampling program)
Start condition:
• Immediate
• Date/time
• Immediate activation
• Time-paced, volume-paced or flow-
paced (CTCV, VTCV, CTVV), external
signal,
• Bottle change after time or number of
samples, external signal
• Bottle synchronization
• Multiple bottles
Stop condition:
• Program end
• Continuous
operation