Liquistation CSF34 Commissioning
Endress+Hauser 55
7.3.3 User definable screens
Menu/Operation/User definable screens
Function Options Info
Meas. screen 1 ... 6 You can create 6 measuring screens of your
own and give them a name. The functions are
identical for all 6 measuring screens.
Meas. screen Selection
• On
• Off
Factory setting
Off
Once you have defined your own measuring
screen, you can switch it on here. You can find
the new screen under User definable screens.
Label Customized text, 20 characters Name of the measuring screen
Appears in the status bar of the display.
Number of lines 1 to 8
Factory setting
8
Specify the number of measured values
displayed.
Line 1 ... 8 User interface
Label
Specify the content of Label in the submenu of
each line.
Source of data Selection
• None
• See list in "Info" column
Factory setting
None
‣
Select a source of data.
You can choose from the following:
• Sensor inputs
• Heartbeat diagnostics of sensor inputs
• Binary inputs
• Current inputs
• Temperature
• Memosens sensor input (optional)
• Fieldbus signals
• Mathematical functions
• Binary inputs and outputs
• Current outputs
• Relay
• Measuring range switching
Measured value
Source of data is an input
Selection
Depends on the input
Factory setting
None
You can display different main, secondary and
raw measured values depending on the type of
input.
No options can be selected for outputs here.
Label Customized text, 20 characters User-defined name for the parameter to be
displayed
Set label to "%0V"
1)
Action If you perform this action you accept the
parameter name that is automatically
suggested. Your own parameter name (Label)
is lost!
1) "%0V" here stands for text that depends on the context. This text is generated automatically by the software and
inserted in place of %0V. In the simplest situations, the generated text could be the name of the measuring
channel, for example.