Micropilot FMR51, FMR52
Endress+Hauser 29
"Power Supply, Output"
1)
"Approval"
2)
Terminal voltage U at the
device
Maximum load R, depending on the supply voltage U
0
at the
supply unit
C: 2-wire; 4-20mA HART,
4-20mA
any 13 to 28 V
3)
R [ ]W
U
0
[V]
10
13 24
20 28
0
500
A0034841
1) Feature 020 of the product structure
2) Feature 010 of the product structure
3) For ambient temperatures T
a
≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).
Polarity reversal
protection
Yes
Admissible residual ripple
at f = 0 to 100 Hz
U
SS
< 1 V
Admissible residual ripple
at f = 100 to 10000 Hz
U
SS
< 10 mV