EasyManua.ls Logo

Endress+Hauser Micropilot FMR51 - Integration into a FOUNDATION Fieldbus Network; Integration into the FOUNDATION Fieldbus Network

Endress+Hauser Micropilot FMR51
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR51, FMR52
Endress+Hauser 61
p
-196
(-321)
-100
(-148)
+310
(+590)
+350
(+662)
+400
(+752)
+450
(+842)
0
(+32)
T
p
[°C]
([°F])
160 (2320)
154 (2233)
148 (2146)
142 (2059)
[bar] ([psi])
A0020555-EN
 18 FMR51: Admissible range of process temperature and process pressure of the HT version (Feature 090
"Seal", Option D2)
Feature 090 "Seal" Admissible process
temperature
Admissible process pressure
A5: Viton GLT –40 to +150 °C
(–40 to 302 °F)
p
rel
= –1 to 40 bar (–14.5 to 580 psi)
C1: Kalrez –20 to +150 °C
(–4 to 302 °F)
D3: Graphite,
40...250°C/-40...482°F
–40 to +250 °C
(–40 to 482 °F)
D2: Graphite,
196...450°C/-321...842°F (HT)
–196 to +450 °C
(–321 to 842 °F)
p
rel
= –1 to 160 bar (–14.5 to 2 320 psi)
Observe the temperature-dependent
restrictions indicated in the diagram
above
1)
.
1) For devices with CRN approval the pressure range may be reduced →  92
FMR52
p
-40
(-40)
0
(+32)
T
p
[°C]
([°F])
+200
(+392)
+100
(+212)
2
1
[bar] ([psi])
25 (362.5)
-1 (-14.5)
0 (0)
A0023250-EN
 19 FMR52: Admissible range of process temperature and process pressure
1 Valid for the following process connections: 6" (ANSI), DN150 (EN), 10K150A (JIS)
2 Valid for all other types of process connection