Loading...
Micropilot FMR60/62/67
26 Endress+Hauser
Repairs
Repair means restoring functional integrity by replacing defective components. Components of
the same type must be used for this purpose. We recommend documenting the repair. This
includes specifying the device serial number, the repair date, the type of repair and the
individual who performed the repair.
The following components may be replaced by the customer's technical staff if genuine spare parts
are used and the appropriate installation instructions are followed:
Component Checking the device after repair
Antennae Visual inspection to check whether all parts are present and properly
mounted
Interference echo mapping must be recorded again
Proof check; test sequence A or B
Seal between antenna and
housing
Visual inspection to check whether all parts are present and properly
mounted
Proof check; test sequence A or B
Display SD02/SD03 Visual inspection to check whether all parts are present and properly mounted
Spare parts kit FHX50 Visual inspection to check whether all parts are present and properly mounted
Main electronics Visual inspection to check whether all parts are present and properly
mounted
Interference echo mapping must be recorded again
Proof check; test sequence A or B
I/O-Module Visual inspection to check whether all parts are present and properly
mounted
Check the measurement at an arbitrary level
Overvoltage protection
OVP10/20
Visual inspection to check whether all parts are present and properly
mounted
Check the measurement at an arbitrary level
Terminals for I/O modules Visual inspection to check whether all parts are present and properly
mounted
Check the measurement at an arbitrary level
Housing cover Visual inspection to check whether all parts are present and properly mounted
Housing cover seals Visual inspection to check whether all parts are present and properly mounted
Housing filters
(vent plugs)
Visual inspection to check whether all parts are present and properly mounted
Safety clamps, housing Visual inspection to check whether all parts are present and properly mounted
Protective cover Visual inspection to check whether all parts are present and properly mounted
Mounting bracket Visual inspection to check whether all parts are present and properly mounted
Slip-on flange Visual inspection to check whether all parts are present and properly
mounted
Proof check; test sequence A or B
Transmitter electronics of the
remote display FHX50
Visual inspection to check whether all parts are present and properly mounted
Cable of the remote display
FHX50
Visual inspection to check whether all parts are present and properly mounted
Protective cover Visual inspection to check whether all parts are present and properly mounted
Mounting bracket Visual inspection to check whether all parts are present and properly mounted
Installation Instructions, see the Download Area at www.endress.com.
The replaced component must be sent to Endress+Hauser for the purpose of fault analysis if the
device has been operated in a protective system and a device error cannot be ruled out. In this case,
always enclose the "Declaration of Hazardous Material and Decontamination" with the note "Used as
SIL device in protective system" when returning the defective device. Please refer to the "Return"
section in the Operating Instructions (→  7).