Micropilot FMR60/62/67
8 Endress+Hauser
Permitted devices types
The details pertaining to functional safety in this manual relate to the device versions listed below
and are valid as of the specified software and hardware version. Unless otherwise specified, all
subsequent versions can also be used for safety functions. A modification process according to
IEC 61508 is applied for device changes.
Valid device versions for safety-related use:
Ordering feature Designation Option
010 Approval All
020 Power Supply;
Output
• A
• B
1)
• C
2)
• K
2-wire; 4-20 mA HART
2-wire; 4-20 mA HART, switch output
2-wire; 4-20 mA HART, 4-20 mA
4-wire 90-253 VAC; 4-20 mA HART
030 Display; operation All
040 Housing All
050 Electrical connection All
070 Antenna All
090 Seal All
100 Process connection All
500 Additional
Operation Language
All
540 Application package All
550 Calibration All
570 Service All
580 Test; Certificate All
590 Additional approval LA
3)
SIL
610 Accessory mounted All
620 Accessory enclosed All
850 Firmware version If no version is selected here, the latest SIL-enabled SW is supplied.
Alternatively, the following SW version may be selected :
72 01.00.zz, HART 7, DevRev01
1) For this version with one current output and one switching output, only the current output (terminals 1
and 2) is suitable for safety functions. The switching output can, if necessary, be wired for non-safety-
oriented purposes.
2) For this version with 2 current outputs, only the first output (terminals 1 and 2) is suitable for safety
functions. The second output can, if necessary, be wired for non-safety-oriented purposes.
3) An additional selection of any further versions is possible.
• Valid firmware version: as of 01.00.zz (→ nameplate of the device)
• Valid hardware version (electronics): as of date of production 28.02.2017 (→ nameplate of the
device)
SIL label on the nameplate
SIL certified devices are marked with the following symbol on the nameplate: