Oxymax COS22 Installation
Endress+Hauser 9
5 Installation
5.1 Installation conditions
5.1.1 Orientation
10° 10°
Not permissible!
Not permissible!
Permissible angle of installation
A0005584-EN
2 Permitted orientations
The sensor must be installed at an angle of inclination of 10 to 170 ° in an assembly,
bracket or appropriate process connection. Recommended angle: 45°, to prevent the
formation of air bubbles.
Inclination angles other than those mentioned are not permitted. Do not install the sensor
overhead.
Observe the instructions for installing sensors in the Operating Instructions for the
assembly used.
5.1.2 Mounting location
• Choose a mounting location that can be easily accessed at a later stage.
• Ensure that upright posts and fittings are fully secured and vibration-free.
• Select an installation location which produces a typical oxygen concentration for the
individual application.