Soliphant T FTM20, FTM21
Endress+Hauser 7
Operating conditions
Installation instructions Mounting location
e.g. storage or buffer container
Orientation
L00-FTM20xxx-11-05-xx-xx-000
Horizontal installation/vertical installation
* With protective cover (to be provided by customer)
** With protecting tube (to be provided by customer)
Environment
Ambient temperature range –40...70 °C (-40...158 °F)
Storage temperature –40...85 °C (-40...185 °F)
Climate class Climatic protection as per DIN IEC 68 Part 2-38, Fig. 2a
Degree of protection IP66/IP67, NEMA4X
Vibration resistance DIN 60068-2-27 / IEC 68-2-27: shock 30 g; vibration 0.01 g
/Hz
Electrical safety IEC 61010, CSA 1010.1-92, FM3600
Electromagnetic compatibility Interference emission to EN 61326, Electrical Equipment Class B
Interference immunity to EN 61326, Annex A (Industrial)
Installation height as per
IEC61010-1 Ed.3
Up to 2000 m (6600 ft) above sea level.