28
2. SET NEXT BUFFER: Selecting this item will display an index of the
ENERAC’s 100 internal storage buffers. Data is automatically stored in the
first available buffer. An asterisk (*) denotes the next storage buffer. If you
want to store data in a different location, use the UP, DOWN, & ENTER
keys to select a new buffer.
3. PRINT BUFFER: This option is used to print data
stored in the analyzer’s memory. Each line
corresponds to one storage buffer. Buffers
containing data show the date (mm/dd) and time
(hh/mm) at which the data was stored. Empty
buffers show the word “empty”. An asterisk (*)
denotes the next storage buffer. To print the
contents of a specific buffer, use the UP &
DOWN keys to move the arrow to select the
desired buffer and press ENTER. If you wish to print all of the ENERAC’s
stored data in sequence, move the arrow to the entry ALL BUFFERS and
press ENTER.
4. ERASE BUFFER: This option is used to erase stored data and is similar to
the PRINT BUFFER MENU. Data that have been stored in the analyzer’s
memory will be retained even after the instrument has been shut off and its
batteries removed. To erase the contents of a specific buffer, use the UP &
DOWN keys to move the arrow to the desired buffer and press ENTER. The
contents of that buffer will be permanently erased. If you wish to erase all
100 of the ENERAC’s stored data, move the arrow to the entry ALL
BUFFERS and press ENTER.
D. PRINTER SETUP MENU
The Enerac 500 has two modes of
printing. In the first mode, the Enerac
will print a record of all parameters as
shown on page 12. In the second mode,
pressing the PRINT key starts a log of
five parameter`s, as shown at right.
Print mode selection (RECORD or LOG), log interval (5 - 240
seconds) and log parameters (COLUMNS 1 - 5) are set in the
PRINTER SETUP MENU.
**PRINTER MENU**
Print: LOG
Interval: 10s
Column #1: ST
ENERAC 500
Serial #:000000
TEST LOG
CUSTOMER NAME
Time: 10:25:00
Date: 07/11/00
Fuel: #2 OIL
ST OXY CO AIR EFF
450 6.0 505 37 80.3
440 6.5 485 43 80.5
425 7.0 460 48 81.0
F % PPM % %
Oxygen Ref:TRUE
** PRINT MODE **
01: 3/31 12:00
02: 4/01 15:30
03:*empty
04: empty
................
99: empty
LL BUFFERS