34
THE ENERAC 500 COMMAND SET
DATA COMMANDS
COMMAND FUNCTION
ATEM? ENERAC returns present value of ambient temperature.
BATT? ENERAC returns the battery voltage.
CDOX? ENERAC returns present value of carbon dioxide.
CMNX? ENERAC returns present value of carbon monoxide.
COMB? ENERAC returns present value of combustible gases.
DRAF? ENERAC returns present value of stack draft.
EFFI? ENERAC returns present value of combustion efficiency.
EXAR? ENERAC returns present value of excess air.
NOXY? ENERAC returns present value of nitric oxide (NO).
NO2Y? ENERAC returns present value of nitrogen dioxide (NO2).
NOXX? ENERAC returns present value of oxides of nitrogen (NOX).
OXYG? ENERAC returns the present value of oxygen.
SO2X? ENERAC returns present value of sulfur dioxide.
STEM? ENERAC returns present value of the stack temperature
TEXT? ENERAC returns a complete record of all current stack
parameters.
SETUP COMMANDS
COMMAND FUNCTION
ATOF? ENERAC returns the ambient temperature offset in °C.
ATOF XX ENERAC sets the ambient temperature offset to XX°C.
COOL? ENERAC returns the thermoelectric cooler duty cycle.
COOL XX ENERAC sets the thermoelectric cooler duty cycle:
XX=50 50% power
XX=100 100% power
CORF? ENERAC returns the temperature units.
CORF X ENERAC sets the temperature units:
X=F Fahrenheit
X=C Celsius
CUST? ENERAC returns the customer name. This name appears on the
display and all printouts.
CUST XXXX ENERAC sets the customer name, up to 21 characters long.
DATE? ENERAC returns the present date.
DATE XX/XX/XX ENERAC sets the present date.