FUEL? ENERAC returns the current fuel used.
FUEL NN? ENERAC returns the fuel currently stored in location #NN.
FUEL NN ENERAC changes its current fuel to fuel #NN (1-15).
MODE? ENERAC returns the current emissions units.
MODE X (Emissions option). Causes ENERAC to switch units of
emissions gas measurements (CO, NO, NO2, NOX, SO2) as
follows:
X=P PPM (volumetric)
X=M MGM (milligrams/cubic meter)
X=# #/B (Lbs./million BTU)
X=G GBH (grams/brake hp-hour)
OXRF? ENERAC returns the oxygen reference.
OXRF XX (Emissions option). Causes ENERAC to set the oxygen correction
factor to any number as follows:
XX=0-20 Percent, in 1% steps
XX=21 TRUE (No correction for oxygen)
PUMP? ENERAC returns pump status: ON, or OFF, and pump duty-
cycle: 0-100%
PUMP0 Turns the sample pump off & turns the purge pump on.
PUMP XX Sets the sample pump duty cycle. (10 < XX < 100)
TIME? ENERAC returns the current time.
TIME XX:XX:XX ENERAC sets the current time. (24-hour format)
MEMORY COMMANDS
COMMAND FUNCTION
BUFF? ENERAC returns the names of each of the storage buffers.
BUFF NNN? ENERAC returns the name of buffer #NN.
BUFF NNN XXXX Sets the name of buffer #NNN to XXXX. Buffer Name can be up
to 11characters.
NBUF? ENERAC returns the total number of storage buffers.
PRNT XXXX Sends to the ENERAC printer the message "XXXX" up to 40
characters long. To send more characters, repeat the command.
PRNT TEXT Commands the ENERAC to print on its printer all the current
stack parameters including time, date, fuel and oxygen reference.
DUMP? ENERAC returns results of all tests stored in its memory.
DUMP NNN? ENERAC returns results of test #NNN.
ERAS NN ENERAC erases the contents of buffer #NNN.
ERAS ALL ENERAC erases the contents of all buffers.