EasyManuals Logo

Engineered air EngA G-TRAC2 User Manual

Engineered air EngA G-TRAC2
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
A G-TRAC2
IOM-25 31 of 42 Mar 99 R7
Johnson Set-Point
4.1 Adjust DSP CAL until the SPC voltage reading matches the dial set point value divided by 10° will give VDC. Return
the dipswitches to their normal position.
Built In Set-Point
4.2 Loosen the setscrew on the set-point knob and move the knob until the dial set point agrees with the measured
set-point value. Tighten the setscrew and return the dipswitches to their normal position.
Room Sensor Calibration
1. Measure the room or return air temperature and set the room set-point dial to the measured value.
2. Record the settings of dipswitches 1, 2, and 3. Turn off dipswitches 1, 2, and 3.
3. Measure the G-TRAC2 calculated set point. (DC voltage from spade connector terminals “DTEMP” to “COMMON”.)
4. Turn dip switch 3 on.
5. Measure the G-TRAC2 calculated set point DC voltage again. Adjust the ROOM CAL pot until the “SPC” to “COMMON”
voltage is the same as was measured in step C.
6. Return the dipswitches to their normal position.
BMS Calibration and Set Up
Refer to notes for BMS (Section XVII, page 23).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Engineered air EngA G-TRAC2 and is the answer not in the manual?

Engineered air EngA G-TRAC2 Specifications

General IconGeneral
BrandEngineered air
ModelEngA G-TRAC2
CategoryController
LanguageEnglish

Related product manuals