Operation
Centrifuge 5424/5424 R
English (EN)
28
To keep the weight differences between the filled tubes low, we recommend taring with a balance. This
will reduce wear on the drive and reduce operating noise.
4. Attach and tighten rotor lid.
5.4.4 Closing the centrifuge lid
1. Check correct attachment of rotor and rotor lid.
2. Push down the centrifuge lid until the lid latch engages and the lid is automatically closed.
The centrifuge will close automatically.
The open key lights up blue. The display shows the symbol.
5.5 Cooling (only 5424 R)
5.5.1 Temperature adjustment
Set the temperature using the temp arrow keys from -10°C to +40°C.
The temperature can also be changed during centrifugation.
With the rotors FA-45-24-11 and FA-45-24-11-Special centrifugation is also possible without a
rotor lid.
Please also note:
The tube lids must be closed securely.
The rotors are not aerosol-tight without rotor lid.
The centrifugation is slightly louder.
Spin columns must always be centrifuged with a rotor lid.
Spin columns
For centrifuging spin columns in the rotor F-45-18-11-Kit, you can leave the tube lids open.
However, this is only permitted with the tubes specified by the kit manufacturers. For reliable
centrifugation, you must lean the open tube lids against the edge of the rotor. Ensure that the
lids do not protrude past the edge of the rotor in the process, then attach the matching rotor
lid.
WARNING! Risk of injury when opening or closing the centrifuge lid.
There is a risk of crushing your fingers when opening or closing the centrifuge lid.
When opening or closing the centrifuge lid, do not reach between the lid and device or into
the latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.