EasyManua.ls Logo

Eppendorf 5424 - Page 8

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Safety
Centrifuge 5424/5424 R
English (EN)
8
2.3.2 Maximum service life for accessories
The rotors listed below, and the corresponding buckets and rotor lids, have a maximum service life of the
number of years or cycles listed in the table (whichever comes first), starting with the initial start-up.
For the other rotors and rotor lids of this centrifuge (see Rotors on p. 18) there is no limit on the service life
as long as the following prerequisites are met: proper use, recommended maintenance and undamaged
condition. To guarantee aerosol tightness, lids of aerosol-tight rotors must be replaced after 50 autoclaving
cycles.
The following requirements must be met in order to use rotors, lids and accessories:
•Proper use,
Recommended maintenance
Undamaged condition
The date of manufacture is stamped on the rotors in the format 03/10 (= March 2010) or on the inside of the
plastic rotor lids in the form of a clock . This is for information only and does not have any reference to
the service life.
WARNING! Risk of injury from chemically or mechanically damaged accessories.
Even minor scratches and cracks can lead to serious internal material damage.
Protect all accessory parts from mechanical damage.
Inspect the accessories for damage before each use. Replace any damaged accessories.
Do not use rotors or buckets with signs of corrosion or mechanical damage
(e.g.,deformations).
Do not use any accessories whose maximum service life has been exceeded.
CAUTION! Risk of injury due to chemically damaged rotor lids or caps.
Transparent rotor lids or caps made from PC, PP or PEI may lose their strength under the
impact of organic solvents (e.g., phenol, chloroform).
If rotor lids or caps have come into contact with organic solvents, they should be cleaned
immediately.
Regularly check the rotor lids and caps for damages and cracks.
Immediately replace rotor lids or caps that have cracks or milky stains.
Accessories Maximum service from the first
commissioning onward
Transparent polypropylene (PP) rotor lids 3 years
Plastic adapters 1 year

Related product manuals