The fixed-angle rotor F-35-30-17 can be loaded with 15 ml Falcons
®
and 15 ml round bottom
centrifuge tubes.
Before centrifuging the Falcons
®
, place their plastic adapters in the tube holders. The adapters
prevent the Falcons
®
from being destroyed. For the centrifugation of round bottom tubes made
of glass, round rubber pads are required to prevent the tubes from cracking. Make sure a rubber
pad is placed under each glass tube before centrifuging.
The maximum load capacity for Falcons
®
is 20 x 15 ml; for 15 ml round bottom centrifuge tubes
30 x 15 ml. The rotor can be loaded with a combination of 15 ml Falcons
®
and 15 ml round
bottom centrifuge tubes. The maximum capacity is then 20 x 15 ml Falcons
®
and
10 x 15 ml round bottom centrifuge tubes
.
Fig. 1:
Loading of rotor F-35-30-17 with 15 ml Falcons
®
and 15 ml round bottom centrifuge tubes
The specified weight imprinted on the rotor is the gross weight rating
including the tube holder, adapter, tubes and contents.
Always weigh the tube holder and the load together.
3.4 Fixed-angle rotor
15 ml Falcon
®
15 ml DIN vessel
3 Operation
Operation
3