Centrifuge 5804/5804 R and 5810/5810 R
8 Technical data
8-4 Technical data
Certificate for the Specifications of Eppendorf Centrifuge 5810
Rotor: A-4-62
Tube: 4 x approx. 50 ml
Rotational speed: 4.000 rpm Tolerance: ± 100 rpm
Run time: 5 min Tolerance: ± 2 s
Measurement Procedure:
Rotational speed measurement: Load the rotor with the specied tubes (tubes should be lled with
ethylene glycol).
Allow the centrifuge to accelerate to the specied rotational speed.
After ve minutes, check the rotational speed.
Measuring instrument used: Stroboscope (accuracy:
1 %).
Time measurement: Set the run time to continuous operation. Wait for ve seconds. Start
the centrifuge by pressing the Start key.
Measure the time which elapses until the display switches from 4 min
to 5 min.
Measuring instrument used: Stopwatch (accuracy: < = 0.1 s)
We herewith certify that each centrifuge is tested according to the above procedures and meets the given
specications.
Hamburg, 15.05.2001
Eppendorf AG
MSTM/KW/0705