Descriptions 10-7
Centrifuge 5804/5804 R and 5810/5810 R
10 Descriptions
PC interface
The centrifuge series 58XX can be equipped with a standard serial interface RS232C and with an
Eppendorf specific interface (BBUS).
For this it is necessary to retrofit the module "IFACE" on the control PCB and to retrofit the 9-pole cable
in the device with SUB_D connector (9-pole). Then it is possible to communicate directly with the
centrifuge via a PC.
A special PC program is also required.
X15 is the plug connection and X8 the plug for the SUB_D cable.
The selection of the interface type is done by PORT 3.5.
PORT 3.5 (SELECT) means "HIGH" RS232C active
PORT 3.5 (SELECT) means "LOW" BBUS active
Pin assignment on the interface cable:
9-pole male SUB_D
1
TX
n
2
RX
n
3 RS232C
n
4
GND
n
5
n
6
V_BUS
n
7
I/O_BUS
n
8 BBUS
GND_BUS
n
9
x
MSTM/KW/0408