Safety
Concentrator plus/Vacufuge® plus
English (EN)
22
3.5.5 Vacuum
WARNING! Risk of injury from direct contact with the vacuum.
Do not expose any body parts to the vacuum of the device.
WARNING! Risk of injury from overpressure.
An overpressure in the exhaust gas lines (e.g. caused by closed taps or blocked pipes) can
lead to bursting.
Only use pipes with a sufficiently large cross section.
Keep the exhaust gas line open.
Do not place any objects on the exhaust gas lines.
Do not flex the exhaust gas lines.
Do not fit any hose clamps or valves to the exhaust gas lines.
Note the maximum pressures and differential pressures permitted (see p. 56).
WARNING! Risk to health from escaping substances.
No vapors from toxic liquids and pathogenic germs must escape.
Ensure the required condensation and separation of vapors using suitable cold traps or
chemical traps.
Please note the personal protective equipment (gloves, clothing, goggles, etc.), extraction
and safety class required for the laboratory.