EasyManua.ls Logo

Eppendorf Mastercycler Gradient - Temperature Increment and Time Increment; Ramp and Ramp Increment Settings

Default Icon
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
7 Programming
+ 0.0
°
Temperature increment (
±
°
C): For each cycle, the temperature is increased or decreased by the
value entered here.
Please note that the temperature cannot exceed 99
°
C. For example, with a temperature
increment of +0.1
°
C and 25 cycles, the starting temperature cannot be greater than 96.5
°
C.
Entry for temperature increase "+"
Entry for temperature decrease " – "
Permitted values
±
0.0 to
±
10.0
°
C
Entry increments 0.1
°
C
+ 0:00
Time increment (
±
s): For each cycle, the time is extended or reduced by the value entered here.
Entry for time extension "+"
Entry for time reduction " – "
Permitted values 0:01 to 1:00
Entry increments 1 s
R = 3
°
/s
Ramp (
°
C/s): The ramp states the speed of the heating or cooling curve of a cycler. The higher
the value entered, the more rapidly the heating or cooling takes place. Most reactions require
rapid heating or cooling rates, i.e. a high value. For processes in which heating or cooling should
take place slowly, different values may be entered for the ramp.
Permitted values 0.3 to 3
°
C/s
Entry increments 0.1
°
C/s
+ 0.0
°
/s
Ramp increment (
±
s): For each cycle, the ramp is raised or lowered by the value entered here.
Please note that the ramp cannot exceed 3
°
C/s, i.e. with an initial ramp entry of 0.3 and a ramp
increment of 0.1, a maximum of 27 cycles is possible.
Permitted values
±
0.1 to
±
1.0
°
C/s
Entry increments 0.1
°
C/s
Possible with Mastercycler
®
gradient only.
G = 0.0
Gradient: When an entry is made here, each column of the block is set to a different temperature.
The nominal temperature set at "T" is in the center of the block. The lowest temperature is on the
left-hand side and increases gradually up to the highest temperature on the right-hand side. This
means that a temperature difference of up to 20
°
C across the block can be generated (see
Sec. 5.4.3). The temperature distribution can be viewed in the OPTIONS/Gradient program. With
0.2 ml test tubes, a maximum of 12 different temperatures may be set whereas only 11 can be
set when 0.5 ml test tubes are used.
Please note that the temperature cannot exceed 99
°
C, i.e. if a gradient of +10
°
C is built up, the
nominal temperature may not be greater than 89
°
C.
Permitted values
±
10
°
C
Entry increments 0.1
°
C
The gradient has no effect when the
"in situ"
type is used.

Related product manuals