System management
Mastercycler
®
nexus
English (EN)
72
System Configuration: Set
device-dependent entries for
ramp and gradient
Making changes under System Configuration is only necessary if you
operate different variants of master and eco devices together.
Mark the System Configuration node. Press the Set softkey.
Set entries for the ramp.
•For devices with an aluminum thermoblock, select Mastercycler
nexus. Adjustable value range: 0.1 - 1.5 °C/s or max..
•For devices with a silver thermoblock, select Mastercycler nexus
SX1. Adjustable value range: 0.1 - 3.0 °C/s or max..
Set entries for the gradient spread.
•Select the setting 20°C for devices with the 96-well thermoblock,
which have a gradient function.If the setting Off is selected, no
gradients can be programmed and the corresponding programs
cannot be started.
•Select the setting 12°C for devices with the 64+32-well thermoblock,
which have a gradient function. If the setting Off is selected, no
gradients can be programmed and the corresponding programs
cannot be started.
Use the Show well types checkboxes to define which well types are
displayed under Edit Grad. (see p. 55).
User Log File: Display or export
user data
Use the User Log File node to export or display all performed actions as a
log file. The following data can be displayed or exported:
•Login and logout procedures
•Changes to and interventions in running programs
•Update of the software version
•Creating and deactivating service intervals
Mark the User Log File node. Press the Show softkey.
The next window enables you to print or export the log file.
Press the Export softkey to export the log file.
Press the Print softkey to print the log file.