Do you have a question about the ERBA ELite 3 and is the answer not in the manual?
Sample positions | 40 |
---|---|
Photometric range | 0 - 3.0 Abs |
Measurement Range | 0 - 3.0 Abs |
Battery | Not applicable |
Sample Type | Serum, Plasma, Urine |
Throughput | 200 tests/hour |
Sample Volume | 2 - 50 μL |
Display | 7-inch color touchscreen |
Connectivity | RS-232, USB |
Operating Temperature | 15-30 °C |
Hematology analyzer for in vitro diagnostic use, developed for small clinics and point-of-care lab offices.
Automated, bench-top hematology cell counter using Coulter-method for cell counting and HGB measurement.
Details the stages of blood testing: aspiration, dilution, WBC measurement, HGB, RBC, PLT analysis.
Details sample volume, chambers, reagent system, throughput, accuracy, display, and power requirements.
Instructions for installation to ensure proper operation and service. Handle with care.
Operate within 15-30°C and 45-85% humidity. Avoid direct sunlight and interfering devices.
Carefully unpack, inspect for damage, check accessories, and acclimate to room temperature.
Turn on via rear switch. Status LED indicates operation. Wait 5 mins for optimal temperature.
Use SHUTDOWN from menu to prevent pneumatic failure. Never just flip power switch.
In emergencies (fire, short-circuit), cut power immediately and use fire extinguisher if needed.
Counts and sizes cells by detecting changes in electrical impedance as particles pass through an aperture.
Measures HGB photometrically after lysing RBCs to release hemoglobin.
Lists and describes 20 hematology parameters measured by the analyzer, including WBC, RBC, HGB, etc.
Specifies accuracy within linearity range. Exceeding range may cause errors or require pre-dilution.
Use K3-EDTA for anticoagulation. Fill tubes to 7-8mm. Invert gently 8 times, do not shake.
Start measurements from MEASURE screen, enter sample info. Results are stored automatically.
Summarizes warning flags for WBC/HGB and recommended user actions.
Summarizes warning flags for RBC/PLT and recommended user actions.
Initiates cleaning procedures for fluidics to reduce blank value, removing contamination from tubing, chamber, valves.
Monitors analyzer stability and verifies performance. Recommended for installation, component replacement, or periodic checks.
Traces analyzer operation and reliability over time. Run control sample daily.
Procedure to verify proper operation of essential components. Performed at installation, after component replacement, or long disuse.
Initiate maintenance procedures like blank measurement, cleaning, priming, or draining chamber.