4.4. Absolute and Linearity Ranges of Parameters
The analyzer provides specified accuracy within its linearity range. Beyond this
linearity range, the instrument can display results but accuracy is impaired.
If a value is over the maximum range of guaranteed linearity, the instrument cannot
measure it and the result will be marked with an E (Error) flag.
To measure a sample, whose parameters exceed the maximum value indicated in
the table below, pre-dilution is recommended. See section 5.2.7 of this manual.
Linearity ranges of primary parameters in normal measuring mode:
Parameter Linearity Ranges Maximum Unit
WBC
0...100 150 10
9
cells/liter
RBC
0...15 20 10
12
cells/liter
PLT
0...700 1000 10
9
cells/liter
HGB
0...250 400 g/l
HCT
0...100 - %
MCV
30...150 - fl
MPV
3...30 - fl
Table 2. Linearity ranges of parameters
Linearity ranges for 1:5 pre-dilution mode:
Parameter Linearity Ranges Maximum Unit
WBC
2…200 300 10
9
cells/liter
RBC
1…30 40 10
12
cells/liter
PLT
100…2000 3000 10
9
cells/liter
Table 3. Linearity ranges of Pre-dilution mode