Calibrating the Model 3301C | 43
7.0 Calibrating the Model 3301C
Before connecting any components, follow
the safety information in the ETS-Lindgren
Product Information Bulletin included with
your shipment.
The ETS-Lindgren 3301CAL
Calibration Fixture
incorporates a resistive
T network and a 10-picofarad
capacitor to perform the
Equivalent Capacitive
Substitution Method (ECSM) of
calibration for rod antennas.
The T allows for accurate
reading of the input to the
3301CAL, and the capacitor
feeds the amplifier through the
same impedance as the rod.
Calibration Steps
Depending on the your calibration configuration, see Model 3301C Calibration
With Receiver and Signal Generator on page 46 or Model 3301C Calibration
With Network Analyzer on page 47.
1. Disconnect the Model 3301C Active Monopole Antenna from power.
2. Remove the monopole element.