17
8. Temperature Setup
Use Setup Temperature mode to desired temperature units; º C or º F.
Press
from the measurement screen and then press ENTER when
TEMPERATURE is selected.
Choose º C or º F temperature units of measure.
Automatic Temperature Compensation (ATC) is automatically applied when a
temperature sensor is connected.
Manual Temperature Compensation (MTC) of 25 º C is used if the temperature
sensor is not connected. See Section 9—Temperature Calibration
to adjust the
MTC temperature value.