READY SNAP
®
METHOD VERIFICATION TEST FOR PART NO.
Ready Snap
®
1P
(plastic ampoules)
Total Alkalinity, pH-II, Calcium Hardness,
Copper, Cyanuric Acid, and Phosphate. 480911
Ready Snap
®
2 Ammonia, Arsenic, Iron, and Manganese 480902
Ready Snap
®
3
Red dye for verifying 525nm eXact
®
Photometer calibration
480903
Ready Snap
®
7
Red dye for verifying 570nm eXact
®
Photometer calibration
480907
ABOUT
EXACT IDIP
®
ACCURACY
Combined with your smart device, the eXact
iDip
®
photometer is
designed to test your water for multiple water quality parameters.
Download the free eXact
iDip
®
app and sync to your smart device
running Bluetooth
®
SMART Technology.
All tests have been calibrated using certi ed reference
standards and analytical spectrophotometric methods. The
algorithms in the app re ect the best correlation of the eXact
iDip
®
against the AWWA, USEPA, DIN and ISO reference
test methods. The eXact
iDip
®
photometer has been factory
calibrated and will stay valid because of its exceptional quality.
We are so con dent in the eXact
iDip
®
photometer, we offer an
industry leading 2-year warranty.
We built the eXact
iDip
®
photometer to be easy, accurate and
environmentally friendly. We have achieved this by utilizing our
patented eXact
®
Strip Micro Technology, which uses 60% less
water and chemistry than alternative methods. Instead of using
a 10mL water sample, eXact
®
Strip Micro uses a 4mL water
sample. The accuracy of the meter is maintained by designing
the sample cell with an 11mm path-length.
Ready Snap
®
is a method veri cation solution
with predetermined values to verify the accuracy
of any manufacturer's tests including our eXact
iDip
®
Smart Photometer System.
The easy 3 step procedure (snap, ll, and test)
allows for quick veri cation of test parameters.
Each box contains 10 ampoules of 10mL
solution with no dilution necessary.
y
METHOD VERIFICATION
EXACT IDIP
®
ASSIGNED VALUE FOR READY SNAP
®
3
READY SNAP
®
3
FREE CHLORINE
DESIRED VALUE
FREE CHLORINE
ACCEPTABLE RANGE
Red dye # 505
1.5 ppm 1.46 – 1.57
EXACT IDIP
®
570 ASSIGNED VALUE FOR READY SNAP
®
7
READY SNAP
®
7
FREE CHLORINE
DESIRED VALUE
FREE CHLORINE
ACCEPTABLE RANGE
Red dye # 22515
2.15 ppm 2.10 – 2.22
Note: Values indicated are speci c to eXact iDip
®
readings and current
concentrations as found at time of manufacture.