25
REMOVE STRIPS
Remove 1 eXact
®
Strip Micro Glycine, Part No. 484014 and 1 eXact
®
Strip Micro
Cd (DPD-1), Part No. 486633 from the bottles before beginning the test. Set the strips in a dry,
convenient place and recap the bottles immediately.
TURN METER ON
Press the button to power the meter on; the display will show annunciators, followed by
the current selection. It will then display the last reading.
RINSE & FILL CELL WITH SAMPLE
Rinse the CELL at least 3 times with the water sample you will be testing—rinsing minimizes
the potential for cross–contamination from a previous test. Finally, fi ll CELL to capacity with
the water sample.
SELECT GROUP & MENU
Press and re–press the button to Select Group 6. Press and re–press the button to
select the test parameter Cd.
DIP STRIP & PRESS READ
Dip the Glycine strip into the CELL, and immediately press . This starts a 20 second
countdown timer. During this time, move the strip in a gentle back and forth motion (approx.
2 strokes/sec). Remove and discard the strip after ‘1’ on the display disappears*. The
meter will auto-zero. Get ready to dip eXact
®
Strip Micro Cd (DPD-1) strip.
DIP STRIP & PRESS READ
About 4 seconds after the meter auto-zeroes, “20” will appear on the display. Immediately dip
the CD (DPD–1) strip into the CELL. During this time, move the strip in a gentle back and
forth motion (approx. 2 strokes/sec). Remove and discard the strip after ‘1’ on the display
disappears.* The meter will begin counting up for 100 seconds, at the end of which, the cursor
will move across the display, while the meter prepares to measures the sample. Record the
result displayed (this result is automatically stored in Cd MENU and, if using the eXact iDip
®
app, will be stored in the app’s ‘RESULTS’). After testing is completed, rinse CELL immediately
and clean with the brush.
1
2
3
4
5
6
Chlorine Dioxide test procedure
Follow the steps to ‘SELECT CUSTOMER’ & ‘CONNECT DEVICE VIA BLUETOOTH’ before
proceeding (see page 12).
*For best accuracy, when testing outdoors in sunlight, place cap over cell cover