COMBINED CHLORINE
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
10
This procedure is a continuation from Free Chlorine Test Method
on previous pages.
DO NOT DISCARD SAMPLE FROM FREE
CHLORINE TEST.
REMOVE STRIP
Remove one eXact
®
Strip Micro CL (DPD-3) Combined
Chlorine and set in a dry, convenient place. Replace
cap on bottle.
9
8
ZERO METER
With the Free Chlorine sample still in the CELL,
press
and the LCD Display reads 0.00
PPM
,
indicating the meter is ready for testing.
FILL, DIP, READ
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP