Programming Parameters - Conductivity
The Table below shows the available programming menu items P1 ~ P9. Each parameter is
explained in detail in the subsequent sections.
Parameters Code Selections
P1 Standard solution series selection
s ol
USA (Europe & U.S.A)
CH (China)
P2 Electrode Constant selection
0.1, 1, or 10
P3 Reference Temperature selection
77, 68, and 64
o
F (25,
20,and 18 )
P4
Temperature compensation
coefficient setting
0.00 to 9.99%
P5 Electrode Constant Calibration
P6 Temperature units
P7 Back light display time setting
0-1-3-6min
P8 Auto power off setting
0-10-20min
P9 Restore to default factory settings OFF / ON
Parameter P1 (Conductivity Solution Setting)
1. From the Conductivity mode, press the MODE key, the ‘P1’ icon appears on the LCD.
2. Use the CAL or the M+/RM keys to toggle through the two (2) selections: USA (for use in the
USA or Europe) and CH (for use in China).
3. Momentarily press MODE to move to the next parameter (P2), or press ENTER to return to the
normal measurement mode.
Parameter P2 (Electrode ConstantK Selection)
1. From the P2 menu, use the CAL or the M+/RM keys to select the desired electrode constant
(0.1, 1.0, or 10). The default setting is K=1.0.
2. Momentarily press MODE to move to the next parameter (P3), or press ENTER to return to the
normal measurement mode.