User's Guide
Digital Conductivity / pH Meter
Model EC600
www.burntec.comwww.burntec.com
Question and Answer IconNeed help?

Do you have a question about the Extech Instruments EC600 and is the answer not in the manual?

Extech Instruments EC600 Specifications

General IconGeneral
Power Supply9V Battery
TypeWater Quality Meter
Temperature32 to 122°F (0 to 50°C)
Resolution0.1 µS/cm, 1 µS/cm, 0.01 mS/cm (Conductivity); 1 ppm, 10 ppm (TDS); 0.1°C/°F (Temperature)
Accuracy±2% FS (Conductivity); ±2% FS (TDS); ±0.5°C/±1°F (Temperature)
Temperature CompensationAutomatic
DisplayLCD

Summary

Keypad and Data Management

Store, Recall, and Clear Datalogger Memory

Instructions on how to store, recall, and clear logged measurement data from the meter's memory.

pH Measurement Operations

3-Point pH Calibration

Detailed procedure for performing a three-point calibration for the pH meter using standard buffer solutions.

1-Point and 2-Point pH Calibration

Guidance on performing one or two-point pH calibrations when the expected measurement range is known.

pH Measurement Considerations

Error Messages and Troubleshooting

Details common error messages (ERR-1, ERR-2) and general troubleshooting steps for pH measurements.

Calibration Intervals and Recalibration Conditions

Guidelines on how often to calibrate and specific situations requiring recalibration for accurate pH readings.

Electrode Care and Maintenance

Instructions on proper handling, cleaning, soaking, and storage of the pH electrode to ensure longevity and accuracy.

Conductivity Measurement and Calibration

Conductivity Calibration Procedure

Procedure for calibrating the conductivity meter using standard solutions.

Measuring Conductivity, TDS, Salinity, Resistivity

Steps to measure conductivity, TDS, salinity, or resistivity of a sample solution.

Conductivity Calibration Considerations

Information on calibration solution series, one-point calibration, and electrode constant (K=1 cm⁻¹).

Conductivity Maintenance and Troubleshooting

General Maintenance and Care

Instructions for keeping the conductivity electrode clean and maintaining the meter's performance.

Troubleshooting and Resetting

Guidance on replacing electrodes and resetting the meter to factory defaults for unusual operation.

Specifications and Technical Details

pH Specifications

Details the measuring range, resolution, accuracy, input current, impedance, stability, and compensation range for pH measurements.

mV Specifications

Details the measuring range, resolution, and accuracy for mV measurements.

Conductivity Specifications

Details measuring ranges, TDS, salinity, resistivity, resolution, accuracy, electrode constant, and temperature compensation for conductivity.