EasyManua.ls Logo

Extech Instruments EC600

Extech Instruments EC600
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EC600 V2.0 4/10
16
Specifications
pH Specifications
Measuring range -2.00 to 19.99 pH
Resolution 0.1/0.01 pH
Accuracy Meter only: ±0.01pH; with probe: ±0.02pH
Input current 2×10
-12
A
Input impedance 1×10
12
Ω
Stability ±0.01 pH / 3hours
Temp. Compensation range 32 to 212
o
F (0 to 100 ) automatic (ATC)
mV Specifications
Measuring range (mV/E
H
) -1999 mV to 0 to 1999mV
Resolution 1mV
Accuracy Meter: ±0.1% FS
Conductivity Specifications
Conductivity measuring ranges
0.00 to 19.99 µS/cm
20.0 to 199.9 µS/cm
200 to 1999 µS/cm
2.00 19.99 mS/cm
20.0 to 199.9 mS/cm
TDS measurement range 0 to 100 g/L (Total Dissolved Solids)
Salinity measurement range 0 to 100 ppt
Resistivity 0 to 100 Mohms
Resolution 0.01/0.1/1 µS/cm and 0.01/0.1mS/cm
Accuracy Meter only: ±1% F.S.; with probe: ±2% F.S.
Electrode constant 0.1 / 1 / 10 cm
-1
Reference temperatures 77, 68, and 64
o
F (25, 20,and 18 )
Temp. Compensation range 32 to 122
o
F (0 to 50 ) (automatic)
www.burntec.comwww.burntec.com