15
Conductivity Measurement, Calibration, and Maintenance Considerations
The meter and probe are calibrated before leaving the factory; the user can take
measurements immediately upon receiving the unit
The recommend calibration period is once per month under normal circumstances; It is
necessary to calibrate a newly purchased conductivity electrode or one that has been
in service for a long period of time
Keep the conductivity electrode clean. It’s best to rinse electrodes with distilled water
and air dry
The surface of the supplied conductivity electrode is plated with a layer of metal
platinum (black) in order to lower the electrode polarization and increase the measuring
range. Do not polish the black platinum surface; clean it by stirring in Distilled water. If
excessive organic buildup appears on the black platinum coating clean with hot water
and detergent or with alcohol
Replace the electrode if the above cleaning methods are ineffective
Reset the meter to its factory default settings (Parameter P9) if unusual operation is
noticed. If the restore process does not solve the issue, return the unit for an evaluation
www.burntec.comwww.burntec.com