Cleaning Recommendations
When cleaning the probe, take care not to scratch or damage the sensing surface or the platinized
electrode surfaces.
Contaminant Cleaning Solution Instructions
Water soluble substances Deionized water Soak or scrub with a soft
brush. Recondition in 4 or 7
buffer for 1 hour.
Grease & Oil Warm water and household
detergent
Soak or scrub with a soft
brush, maximum of 10
minutes. Rinse thoroughly
with DI water, recondition in 4
or 7 buffer for 1 hour.
Heavy grease & Oil Alcohol Maximum of 5 minute soak,
scrub with a soft brush. Rinse
thoroughly with DI water,
recondition in 4 or 7 buffer for
1 hour.
Lime and hydroxide coatings 10% acetic acid Soak until coating dissolved,
maximum of 5 minutes. Rinse
thoroughly with DI water,
recondition in 4 or 7 buffer for
1 hour.
Please Note: Since the EC500 does not have a refillable pH reference electrolyte chamber, it is
important not to soak the electrode in the above solutions for more than the recommended times. To
do so may cause a reference potential shift, which will cause degradation in performance or failure.
w ww. . com
information@itm.com1.800.561.8187

Other manuals for Extech Instruments ExStik EC500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Extech Instruments ExStik EC500 and is the answer not in the manual?

Extech Instruments ExStik EC500 Specifications

General IconGeneral
BrandExtech Instruments
ModelExStik EC500
CategoryMultimeter
LanguageEnglish

Related product manuals