Specifications
Electrical Specifications (23±5°C)
Humidity/Air Temperature
Unit Range Resolution Accuracy
%RH 10 to 95 0.1 <70%: ±(4%RH)
≥70%: ±(4%rdg + 1.2%RH)
Temp
o
C 0 to 50
0.1 ±1.2
o
C
Temp
o
F 32 to 122
0.1 ±2.5
o
F
Type K Thermocouple (external probe)
Unit Range Resolution Accuracy*
Type K Temp
o
C
-50.0 to 1300.0 0.1 ±(0.4%rdg + 0.5
o
C)
-50.1 to -100.0 0.1 ±(0.4%rdg + 1.0
o
C)
Type K Temp
o
F
-58.0 to 2372.0 0.1 ±(0.4%rdg + 1.0
o
F)
-58.1 to -148.0 0.1 ±(0.4%rdg + 1.8
o
F)
*Accuracy spec. does not include additional error introduced by the external probe
Dew Point Temperature
o
C -25.3 to 49.0 0.1 Calculation based on
humidity/temp values.
o
F -13.5 to 120.0 0.1
Wet Bulb Temperature
o
C -5.4 to 49.0 0.1 Calculation based on
humidity/temp values.
o
F 22.2 to 120.0 0.1