ThermocoupleInput
(typesK,J,T,E,R,andS)
Thesuppliedthermocoupleprobesareuseableintherangeof‐20to
+250°C(‐4to+482°F)only.For
measuringoutsideofthisrange,pleaseobtainthermocouplesratedaccordingly.
Accuracyspecifiedformeteronly.Inputprobesaddadditionalmeasurementerror.GenerallyRTD
probesintroduceasmallererrorthanthermocouples.
Theabovespecifications
aretestedunderanenvironmentalRFFieldStrengthlowerthan3V/Mand
afrequencylowerthan30MHz
SensorType Resolution Range Accuracy
TypeK
0.1
°
C
‐100.0to‐50.1
°
C
±(0.4%+1
°
C)
‐50.0to 999.9
°
C
±
(0.4%+0.5°
C)
1°C
1000to1300
°
C
±(0.4%+1
°
C)
0.1
°
F
‐148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
‐58.0to999.9
°
F
±(0.4%+1
°
F)
1°
F
1000to2372
°
F
±(0.4%+2
°
F)
TypeJ
0.1
°
C
‐100.0to‐50.1
°
C
±
(0.4% +1°
C
)
‐50.0to 999.9
°
C
±(0.4%+0.5
°
C)
1°
C
1000to1150
°
C
±(0.4%+1
°
C)
0.1
°
F
‐148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
‐58.0to999.9
°
F
±
(0.4% +1°
F
)
1
°
F
1000to2102
°
F
±(0.4%+2
°
F)
TypeT
0.1
°
C
‐100.0to‐50.1
°
C
±(0.4%+1
°
C)
‐50.0to 400.0
°
C
±(0.4%+0.5
°
C)
0.1
°
F
‐148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
‐58.0to752.0
°
F
±
(0.4% +1°
F
)
TypeE
0.1
°
C
‐100.0to‐50.1
°
C
±(0.4%+1
°
C)
‐50.0to 900.0
°
C
±(0.4%+0.5
°
C)
0.1
°
F
‐148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
‐58.0to999.9
°
F
±
(0.4% +1°
F
)
1
°
F
1000to1652
°
F
±(0.4%+2
°
F)
TypeR
1
°
C
0to600
°
C
±(0.5%+3
°
C)
601to1700
°
C
1°
F
32to1112
°
F
±(0.5%+5
°
F)
1113to3092 °F
TypeS
1
°
C
0to600
°
C
±(0.5%+3
°
C)
601to1500 °C
1°
F
32to1112
°
F
±(0.5%+5
°
F)
1113
to2732
°
F