Page #9 background imageLoading...
Page #9 background image
9 TM500-EU-EN v1.1 12/13
DataloggingandPCInterface
TypesofDataRecording
ManualDatalogging:Manuallylogupto99readingsontoanSDcardviapushbuttonpress.
Thesampleratemustbesetto0toimplementthismode.
AutomaticDatalogging:AutomaticallylogdataontoanSDmemorycardwherethenumber
ofdatapointsisvirtuallylimitedonlybythecardsize.Readingsareloggedataratespecified
bytheuserfrom1to3600seconds.
RS232/USB:ThemeterincludesanRS232/USBPCinterfacejacklocatedonthelowerright
sideofthemeterunderthesn
apoffcompartmentcover.ForstreamingofdatatoaPCvia
theRS232Outputjack,theoptional407001USBkit(USBcableanddriverCD)alongwiththe
407001softwareare required.
SDCardInformation
InsertanSDcard(from1Gsizeupto16G)intotheSDcardslotatthebottomofthemeter.
Thecardmustbeinsertedwiththefrontofthecard(labelside)facingtowardtherearofthe
meter.
IftheSDcardisbeingusedforthefirsttimeitisrecommendedthatthecardbeformatted
andthelogger’sclocksettoallowforaccuratedate/timestampingduringdatalogging
sessions.RefertotheSetupModesectionforSDcardformattingandtime/datesetting
instructions.
EuropeanandUSAnumericalformatsdiffer.ThedataontheSDcardcanbe formattedfor
eitherformat.ThemeterdefaultstoUSAmodewhereadecimalpointisusedtoseparate
unitsfromtenths,i.e.20.00.TheEuropeanformatusesacomma,i.e.20,00.Tochangethis
setting,refertotheSetupModesection.
ManualDatalogging
InthemanualmodetheuserpressestheLOGbuttontomanuallylogareadingontotheSDcard.
1. Setthesamplingrateto‘0’secondsasdescribedintheSetupModesection.
2. PressRECbuttontoturnonmanualrecordmode.TheRECiconwillappearintheupper
leftcornerofthedisplay.Thelowerportionofthedisplaywillshowpn(n=memory
positionnumber199).
3. PresstheENTERbutton momentarilytologareadingintomemory.TheLOGGERiconwill
flasheachtimeadatapointisstored(theSCANSDiconwillapp earwhenthe meter
accessesthecard).
4. Advancetothenextmemorylocationusingtheuparrowbutton.
5. Toexitthemanualdataloggingmode,pressandholdtheRECbuttonforatleast1.5
seconds.
netzerotools.com
netzerotools.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Extech Instruments TM500 and is the answer not in the manual?

Extech Instruments TM500 Specifications

General IconGeneral
BrandExtech Instruments
ModelTM500
CategoryData Loggers
LanguageEnglish

Related product manuals