5: Maintenance /Service
47
Specific for FPGA Errors
Analyzer Index
Timeout at initialization while
searching the index of the
abort.
or Contact Falex Rep/Compass
Compensation Error
during compensation is below
will abort.
or Contact Falex Rep/Compass
Edge Detection
During edge detection the tube
translation axis will hit the back
limit switch or the sensor signal
won’t be above the threshold
will abort.
‘Tube Translation’ or Power
cycle the unit or Contact
Falex Rep/Compass Service
Measurement Values
The difference between min and
max value for one measurement
point is smaller than the value
will abort.
Can happen for very thick
Deposits. Power cycle the
Unit or contact Falex Rep/
Shutter not closed
Shutter is not closing during
Tube insert or eject.
‘Shutter’ or Power
cycle the unit or Contact
Falex Rep/Compass Service
Shutter, Limit Switches
Both Limit Switches are triggered
or there is a timeout when
moving the axis.
will abort.
‘Shutter’ or Power
cycle the unit or Contact
Falex Rep/Compass Service
Compensation,
Limit Switches
Both Limit Switches are triggered
or there is a timeout when
moving the axis.
will abort
‘Compensation’ or Power
cycle the unit or Contact
Falex Rep/Compass Service
Tube Rotation, index
Timeout at initialization while
Searching the index of the tube
Rotation axis.
will abort
‘Tube Rotation’ or Power
cycle the unit or Contact
Falex Rep/Compass Service
Tube Translation,
Limit Switches
Both Limit Switches are triggered
or there is a timeout when
moving the axis.
will abort
‘Tube Translation’ or Power
cycle the unit or Contact
Falex Rep/Compass Service
Tube Translation,
Init path
The measured path from one
limit switch to another is not
within the tolerated range
will abort
‘Tube Translation’ or Power
cycle the unit or Contact
Falex Rep/Compass Service