FS10 Series
Fluid Components International LLC 3
Flow Direction Arrow
Sensors perpendicular
“A” (Active) mark
Upstream
Flow Direction
Sensors parallel
Left to Right Flow Example in 1/4 and 3/8 inch tube tees
For optimum sensitivity in low flow applications, install probes
with the “A” (Active) sensor positioned upstream.
Horizontal lines: gas or liquid.
Vertical lines: gas - flow must be down
liquid - flow must be up
FS10i
Install probe with reference flats parallel to flow and
arrow on top in horizontal installations.
Right to Left Flow Example in 1/2 inch or greater tube or
pipe tee
Probes mounted into 1/2 inch or greater tees are installed
with the sensors positioned perpendicular to the flow path in
the tee. Orient with the “A” (Active) facing up in side mount
horizontal configurations. These sensors may also be marked
with a flow direction arrow.
Mounting Orientation
Figure 1 – Mounting Orientation