ST51 MASS FLOW FLUID COMPONENTS INTERNATIONAL LLC
This page is subject to proprietary rights statement on last page
13 Doc. No. 06EN003389 Rev. -
Calibration Parameters Verification
The instrument uses a set of predetermined calibration parameters to process flow signals. Most of these parameters should not
change. A data package located with this manual contains the “ST51 Delta R Data Sheet”. This contains the calibration parameters
stored in the flow transmitter at the factory. To verify that these parameters have not changed, complete the following:
1. Identify the appropriate Delta R Data sheets by serial number of the instrument.
2. Press [D] [ENTER] to examine each of the parameters. The [ENTER] key allows scrolling one message at a time. Use Table
4 to verify parameters with the Delta R Data sheet ST51 Parameters.
Table 4. Diagnostic Test Sequence on Display
S/W Version: dR Min: T SpanIDAC 0:
Flow Factor: dR Max: T ZeroIDAC 0:
Cmin Flow: Cal Ref: T SpanIDAC 1:
Cmax Flow: Tcslp: T ZeroIDAC 1:
Eng Units: Tcslp 0: State 0:
Line Size 0: Tcslp 2: Switch Pt 0:
Line Size 1: Tot Menu: State 1:
Cmin Temp: Tot Flag: Switch Pt 1:
Cmax Temp: Totalizer: K factor 1:
Min Flow: Rollover Cnt: K factor 2:
Max Flow: Fix Pt Flag: K factor 3:
Density: Pulse Factor: K factor 4:
*C1 [1]: Pulse Out: I factor:
*C1 [2]: Hours: Temp Flag:
*C1 [3]: Sample Period: Out Mode:
*C1 [4]: dR Slope: Boxcar Max:
*C1 [5]: dr Off Set: RTD-SLP-385:
Break Pt: Refr Slope: % of Range:
*C2 [1]: Refr Off Set: Us er Name:
*C2 [2]: SpanIDAC 0: Shop Order #:
*C2 [3]: ZeroIDA C 0: Serial No.:
*C2 [4]: SpanIDAC 1: Model#:
*C2 [5]: ZeroIDA C 1:
If parameters that have an asterisk (*) have changed, this may indicate a problem. Customer Service should be contacted. If the
parameters have not changed, continue with the next section.