EasyManua.ls Logo

FGH 2000 Series - Read Messages; Set Instrument Status Messages; Message Header

FGH 2000 Series
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SERIES 2000 COMMUNICATIONS MANUAL 5
M41 Issue 1.1
messages to the instrument may have the fields separated by spaces. These spaces will be
ignored. Messages from the instrument will not contain spaces. Eg.
W 45 C 0123 <CR>
will attempt to write 123 to the local set point as well as
W45C0123<CR>
The <CR> at the end of the message is a carriage return. (Hex 0D). Each message written
to the instrument must be terminated with, and each message from the instrument will be
terminated with a carriage return.
3.1.2 Read messages
Read messages to the instrument take the following form.
RAAPSS<CR>
where
R
= read message header
AA
= instrument address
P
= parameter code to be read
SS
= segment number (P2000 Only)
Again the message must be terminated with a <CR>, and spaces may be included if
desired.
3.1.3 Set instrument status messages
Set messages to the instrument take the following form.
SAAC<CR>
where
S
= set message header
AA
= instrument address
C
= set parameter code to be performed
Again the message must be terminated with a <CR>, and spaces may be included if
desired.
3.2 Message header.
This may be;
ASCII
R
signifying a Read
ASCII
W
signifying a Write
ASCII
S
signifying a Set.
The
R
header is used whenever data is being read from the instrument. When this header
is used the data field is absent. The
W
header is used to write data to the instrument. The
S
header is used to set the status of the instrument.