Standard Incubators
600 Series
Operation Manual and Parts List
LT2068X1 Rev. 2 (7006841)
© 2010 Fisher Scientific Inc.
All rights reserved
Printed in U.S.A.
Isotemp

Questions and Answers

  • S
    Sandra MeyerAug 10, 2025
    What to do if my Fisher Scientific 6845 Laboratory Equipment shows 'EEE' on the display?
    • B
      Brian PerkinsAug 10, 2025
      If the alarm LED is also lit, allow the incubator to cool to the set point. If the alarm LED is not on, check the thermocouple connections at the rear of the temperature controller. If the thermocouple is OK, replace the control.
  • D
    Danny ReedAug 15, 2025
    How to fix a Fisher Scientific 6845 that is not heating?
    • C
      Cassandra JohnstonAug 15, 2025
      First, ensure the set temperature is not less than the actual temperature. If not, replace the control thermocouple. Also, tighten connections at the terminal strip and/or heater. Check the heater resistance against the schematic and replace if necessary. Finally, replace the controller.
  • P
    Peter EscobarAug 21, 2025
    What to do if my Fisher Scientific 6845 incubator temperature is erratically high?
    • V
      victoriagarciaAug 21, 2025
      If the incubator temperature is erratically high, the control thermocouple or the control board may be defective and require replacement.
  • D
    David SmithAug 28, 2025
    Why is the alarm LED on and the control higher than the set temperature on my Fisher Scientific 6845 Laboratory Equipment?
    • S
      Sheila CastroAug 28, 2025
      This can happen if the set temperature has been changed to a value less than the actual temperature minus the alarm limit. Wait for the actual temperature to cool to the set temperature, specifically to set plus 3°, and the LED should go out and the EEE alarm display will return to normal operation. If this doesn't work, the controller or solid state relay/safety relay may be defective and require replacement.
  • D
    Dylan HinesSep 2, 2025
    How to troubleshoot a Fisher Scientific 6845 Laboratory Equipment with no power?
    • J
      Jessica AndersonSep 2, 2025
      First, ensure the unit is plugged in and turned on. If it is, replace the circuit breaker if it's defective, or replace any loose wires.

Summary

Section 2 Specifications

Performance Specifications

Covers operating range, temperature uniformity, resolution, sensitivity, and recovery time for the incubators.

Section 3 Installation

Preparing the Incubator for Operation

Step-by-step guide to prepare the incubator, including installing shelves and connecting power.

Section 4 Controls

Keypad Functions

Explains the function of each key on the incubator's four-key tactile keypad for operation and settings.

Section 5 Operation

Setting the Incubator Temperature

Step-by-step guide on how to set and adjust the incubator's target temperature using the keypad.

Safety Precautions for Operation

Outlines essential safety precautions to follow during incubator operation, including material compatibility and personal protective equipment.

Section 6 Service

Safety, Cleaning, and Gasket Replacement

Covers general safety cautions, cleaning procedures, and replacing the door gasket.

Section 7 Troubleshooting

Section 8 Replacement Parts